Table 1 Precision of VerteTrack applied load

From: The bench-top accuracy of the VerteTrack spinal stiffness assessment device

Indenter head loading

LoadVerteTracka (N)

95%CI (N)

SD

CV

RIH only

16.557

16.470 to 16.645

0.045

0.27%

RIH + 1 plate

27.757

27.701 to 27.814

0.029

0.10%

RIH + 2 plates

38.662

38.589 to 38.735

0.037

0.10%

RIH + 3 plates

49.583

49.479 to 49.688

0.053

0.11%

RIH + 4 plates

60.687

60.592 to 60.783

0.049

0.08%

RIH + 5 plates

71.461

71.420 to 71.502

0.021

0.03%

  1. CV Coefficient of variation, RIH Rolling indenter head, SD standard deviation
  2. aAverage of 10 measurements at each load. All loads measured with digital scale (OHAUS, model TS4KD: Resolution = 0.1 g, accuracy ±0.07 g. Equivalent to resolution = 0.001 N, accuracy ±0.0007 N)